Cannon Design Product Group


             
     

eVap 1000

eVap 1100

eVap 1500

 
 
Dimensions 10.5"w x 7"d x10.5"h
(26.7 cm x 17.8 cm x 26.7 cm)
10.5"w x 7"d x10.5"h
(26.7 cm x 17.8 cm x 26.7 cm)
34"w x 16"d x 25"h
(86 cm x 41 cm x 64 cm)
Hose Connections (Lab Water and Building Water) 3/8" (9.5 mm) 3/8" (9.5 mm) 1" (25.4 mm)
Nominal Pump Capacity Gpm (lpm) 0.5 (1.9) 3.0 (11.4) 5 gpm (18.9)
Cooling Capacity at 20° F dT, btu/hr (watts) 5,000 (1500) 30,000 (8800) 50,000 (14,700)
Building Water Flow Rate at 10° F dT, gpm (Lpm) 1.0 (3.79) 6.0 (22.80) 10 (37.9)
Pressure, (PSI) (2.0) (2.0) (70)
Shipping Weight, lbs (kg) 13 (6) 14 (6.4) 270 (122)
Speed Variable Variable Variable
Electrical, Maximum Power Consumption 110 V/1 Ph, 12 watts 110 V/1 Ph, 26 watts 110 V/1 Ph, 5 amps
 
             
 

ALL UNITS:

Coolant: Building Chilled Water or Process Water | Operating Temperature Range: 40° to 180° F (4° to 71° C)

 

All units are constructed from lead-free and corrosion-resistant materials including brass, copper and 316 Stainless Steel. All units are portable. The 1000 and the 1100 come with the

standard ability to be bench mounted, wall mounted, or mounted to fume hood lab racks to minimize space requirements. The 1500 comes with digital PLC temperature Control up to

+/- 0.1 degree C, and variable speed control, remote DDC monitoring, conductivity monitor, and locking casters.

 
   
  EVAP 1000/1100 I EVAP 1500 I SPECIFICATIONS I ACCESSORIES